Spiral / Helical CT 2D Low-Contrast Resolution Phantom with two contrast steps
Measure the in-plane and axial low-contrast resolution. Optimize tube current, collimation, pitch and image reconstruction for the desired low-contrast resolution in all types of clinical applications.
Order-number: QRM-2DLC |
download product data sheet: |
![]() |
![]() |
click image to enlarge 
The Phantom has been designed to evaluate the imaging capabilities of a CT scanner in the x/y-plane as well as in the axial-plane. The CT’s low-contrast resolution capabilities can be obtained by a single spiral scan using axial images and coronal reformations. The phantom visualizes the impact of all scan, image reconstruction, and display parameters.
In order to mimic imaging conditions in either head or body applications, a contrast of -10 HU or -20 HU can be ordered. Other contrast levels are available on request.
Two series of different low-contrast cylinders with diameter varying from 5 mm to 15 mm are located in the 100 mm diameter cylindrical body of homogeneous tissue-equivalentmaterial.
Specifications |
|
| Phantom diameter: | 100 mm |
| Phantom length: | 100 mm |
| Phantom weight: | 0.9 kg |
| Material: | tissue-equivalent plastic, typ. 38 HU (at 120 kV) |
| Contrast inserts: | -10 HU and -20 HU relative to background |
| Cylindrical contrast inserts: | 5 mm 10 mm 15 mm |
| Accuracy: | ±3 HU of specified values ±1 HU of certified values |