Micro-CT Low Contrast Phantom
As a useful tool for evaluation of contrast detectability the MicroCT Low Contrast Phantom provides four inserts of two different contrast levels, each in two different dimensions.
Order-number: QRM-MicroCT-LC |
download product data sheet: |
![]() |
![]() |
![]() |
click image to enlarge 
The MicroCT Low Contrast Phantom is a
cylindrical phantom in which small details
(cylindrical inserts) at a specified contrast with
respect to the background allow for an
estimation of the low-contrast resolution of a
Micro-CT scanner. The difference in
Hounsfield Units can be as low as -40 HU and
-80HU, respectively.
Actual values vary slightly depending on
scanner geometry, calibration and tube
voltage.
Standard phantoms are scaled down from the
diameter of an average human trunk (32 cm)
to that of a fat mouse or lean rat (32 mm).
Other dimensions as well as inserts with
substances other than specified here can be
produced upon request.
Specifications |
|
| Base material: | resin |
Low contrast inserts: |
|
| Diameter: | 1 mm / 2.5 mm |
| &Delta HU: | ∼ 4 % / 8 % |
| approx. HU contrast: | -40 / -80 HU |
Dimensions for mouse-size phantom: |
|
| Diameter: | 32 mm |
| Length: | 40 mm |
| Weight: | 35 g |