3D Spatial Resolution Phantom

For the visual evaluation of spatial resolution in CT and Cone-Beam CT

  • 2 different versions available
    • QRM-10111 3D Spatial Resolution Phantom (stand-alone)   
    • QRM-10101 3D Spatial Resolution Phantom, D100 
  • Evaluate spatial resolution in-plane and sagittal CT images
  • Provides several hole patterns in two planes

The high-contrast spatial resolution test phantom visualizes spatial resolution axial and sagittal plane of dedicated CT protocols and image reconstruction algorithms. The test pattern is a series of drilled holes with varying diameter and spacing from 4.0 mm down to 0.4 mm (see Specifications) allowing for an order of magnitude in spatial frequency. With spiral/helical CT, evaluating both axial images and coronal reformations, spatial 3D resolution can be tested by a single scan.

 

Options for QRM-10101: 3D Spatial Resolution Phantom, D100 

Dimension (in-plane): 140 mm x 173 mm
Height (z): 194 mm
Weight: 0.8 kg
Size of plates with test patterns: 50 mm x 100 mm x 10 mm
Body material acrylic plastic
   
Diameter/mm lp/cm
4.0 1.25
3.0 1.66
2.0 2.50
1.5 3.33
1.2 4.16
1.0 5.00
0.9 5.55
0.8 6.25
0.7 7.14
0.6 8.33
0.5 10.0
0.4 12.5
Diameter 100 mm
Height (z): 70 mm
Weight: 0.8 kg
Two plates with test patterns aligned perpendicular:  
  Ø 100 mm x 10 mm
  50 mm x 100 mm x 10 mm
Body material: acrylic plastic
Diameter/mm lp/cm
4.0 1.25
3.0 1.66
2.0 2.50
1.5 3.33
1.2 4.16
1.0 5.00
0.9 5.55
0.8 6.25
0.7 7.14
0.6 8.33
0.5 10.0
0.4 12.5

Related Downloads

Spec
3D Spatial Resolution Phantom ( 2 MB )
Catalog
QRM Phantoms for X-ray Imaging ( 5 MB )
Diese Seite existiert auch in Ihrer Sprache.
This page also exists in your language.
Sprache wechseln Change Language